Voltage and current calibrator FlukeF715 FLUKE NEW

Voltage and current calibrator FlukeF715 FLUKE NEW
Voltage and current calibrator FlukeF715 FLUKE NEW
Voltage and current calibrator FlukeF715 FLUKE NEW
Voltage and current calibrator FlukeF715 FLUKE NEW
Voltage and current calibrator FlukeF715 FLUKE NEW
PLCAGE: Voltage and current calibrator FlukeF715 FLUKE NEWFunctional specificationsMeasurement / Out... Read More
$1,983.00
Ex Tax: $1,983.00
  • Stock: In Stock
  • Model: FlukeF715
  • Weight: 0.35kg
1. We typically ship items within three business days after payment is received.
2. Shipping options include DHL, UPS, FedEx, and EMS.
3. Estimated delivery time for express shipping via DHL, UPS, FedEx, or EMS generally ranges from 3 to 10 business days, depending on the destination country.
4. Please verify your address thoroughly when placing your order.
5. Important for buyers: You are responsible for any tariffs on accessories. To potentially avoid these tariffs, you may choose EMS express shipping.
6. The shipping service is responsible for any accidents, delays, or other issues that may arise.

Returns:
We offer a 14-day return policy. If you wish to return an item, please contact us within 14 days and ensure that the package remains unopened. The buyer is responsible for the shipping costs for both the return and the initial delivery, as well as any applicable customs fees. Once we receive and inspect the product, we will process your refund within 24 hours.
For any quality issues that arise during normal use within the warranty period, we will repair the item free of charge. The buyer is responsible for the return shipping cost, while our company will cover the cost of re-shipping the repaired item.
We accept many payments,include Paypal,Bank Transfer etc
Tags: FlukeF715

PLCAGE: Voltage and current calibrator FlukeF715 FLUKE NEW

Functional specifications

Measurement / Output mV

Range:

0 to 100 mV

Resolution:

0.01 mV

Accuracy:

0.02% + 2 counts

Measurement / Output V

Range:

0 to 10 V

Resolution:

0.001 V

Accuracy:

0.02% + 2 counts

Measuring / output / analog mA (% span)

Range:

0 to 24 mA (-25% to 125%)

Resolution:

0.001 mA

Accuracy:

0.015% + 2 counts

Note: The receiving end (analog transmitter) Display mA or% span

Specification

Loop Power

Range:

24 V dc

Accuracy:

+ / - 10%

Note: Drive: 20 mA to 1000 Ω, for battery> 6.8V; 700 Ω, the battery voltage level is 6.8 V 5.8

Display

5 digit LCD display

Environmental Specifications

Operating temperature

-10 º C to 55 º C

Storage temperature

-40 º C to 60 º C

Relative humidity (non-condensing)

95% (10 º C to 30 º C)

75% (30 º C to 40 º C)

45% (40 º C to 50 º C)

35% (50 º C to 55 º C)

Operating altitude

3,000 m maximum

Safety Specifications

Security

CSA C22.2 No. 1010.1:1992

Mechanical & General Specifications

Size

187 mm L x 87mm W x 32 mm

201 mm L x 98mm W x 52 mm (with protective cover)

Weight

Approximately 330 g

Approximately 600 g (with protective cover)

Warranty

3 years

Shock and vibration

Random, 2G, 5-500 Hz

1m drop test

Battery Type

9V battery, ANSI / NEDA 1604A or IEC 6LR619V alkaline