We use cookies 🍪
We use cookies and other similar technologies to improve your browsing experience and the functionality of our site. Learn more in our Privacy Policy.

AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW

AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW
AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW
AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW
AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW
AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW
AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW
AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW
AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW
AC/DC 2000A CLAMP METER FLUKE F355 FLUKE NEW
PLCAGE: AC/DC 2000A CLAMP METER FLUKEF355 FLUKE NEWFeatures:Reliably handle a wide range of hig... Read More
$1,063.00
Ex Tax: $1,063.00
  • Stock: In Stock
  • Model: FLUKE F355
  • Weight: 2.00kg
1. We typically ship items within three business days after payment is received.
2. Shipping options include DHL, UPS, FedEx, and EMS.
3. Estimated delivery time for express shipping via DHL, UPS, FedEx, or EMS generally ranges from 3 to 10 business days, depending on the destination country.
4. Please verify your address thoroughly when placing your order.
5. Important for buyers: You are responsible for any tariffs on accessories. To potentially avoid these tariffs, you may choose EMS express shipping.
6. The shipping service is responsible for any accidents, delays, or other issues that may arise.

Returns:
We offer a 14-day return policy. If you wish to return an item, please contact us within 14 days and ensure that the package remains unopened. The buyer is responsible for the shipping costs for both the return and the initial delivery, as well as any applicable customs fees. Once we receive and inspect the product, we will process your refund within 24 hours.
For any quality issues that arise during normal use within the warranty period, we will repair the item free of charge. The buyer is responsible for the return shipping cost, while our company will cover the cost of re-shipping the repaired item.
We accept many payments,include Paypal,Bank Transfer etc
Tags: FLUKE F355
PLCAGE: AC/DC 2000A CLAMP METER FLUKEF355 FLUKE NEW

Features:

Reliably handle a wide range of high-current applications with 2000 A ac + dc true-rms, 1400 A ac, and 2000 A dcThe large 58 mm (2.3 in) jaw capacity is suitable for large or multiple conductorsCAT IV 600 V, CAT III 1000 V rating for added user protectionIn-rush current measurement captures ‘power-on’ surge current with accuracy and repeatabilityHigh voltage measurement of 1000 V ac + dc true-rms, 600 V ac, and 1000 V dc allows the user to perform multiple tests with only one tool (355 only)Resistance to 400K ohms coupled with a continuity beeper provide the convenience of a multimeter in a clamp meter. (355 only)Accurately measure frequency up to 1 kHz for optimum troubleshootingQuickly analyze readings using the MIN, MAX, and AVG functionsA large backlit display allows for easy visibility in low-lit areasUse the display hold feature to capture readings even when the display cannot be viewedUse the low-pass filter to smooth out noisy loads and stabilize readings

Electrical specifications

Current measurement dc and ac 10 Hz to 100 Hz

Range : 40 A

 Resolution: 10 mA
Accuracy, A: 1.5 % rdg + 15 digits
Trigger Level for Inrush: 0.50 A
Trigger Level for Hz Filter OFF: 2.50 A
Trigger Level for Hz Filter ON: 0.50 A

Range : 400 A

 Resolution: 100 mA
Accuracy, A: 1.5 % rdg + 5 digits
Trigger Level for Inrush: 5.0 A
Trigger Level for Hz Filter OFF: 2.5 A
Trigger Level for Hz Filter ON: 2.5 A

Range : 2000 A; 1400 ac rms

 Resolution: 1 A
Accuracy, A: 1.5 % rdg + 5 digits
Trigger Level for Inrush: 5 A
Trigger Level for Hz Filter OFF: 8 A
Trigger Level for Hz Filter ON: 8 A

 

Crest Factor (50/60 Hz)

Range : 40 A

 Crest Factor* : 2 @ 33 A, 2.4 @ 27 A

Range : 400 A

 Crest Factor* : 2 @ 330 A, 2.4 @ 270 A

Range : 2000 A; 1400 ac rms

 Crest Factor* : 2 @ 1000 A, 2.4 @ 833 A

 

Current measurement ac 100.1 Hz to 1 kHz

Range : 40 A

 Resolution: 10 mA 
Accuracy > 10 A : 3.5 % rdg + 15 digits
Trigger Level for Inrush: 0.50 A
Trigger Level for Hz Filter OFF : 2.50 A
Trigger Level for Hz Filter ON : 0.50 A

Range : 400 A

 Resolution: 100 mA 
Accuracy > 10 A : 3.5 % rdg + 5 digits
Trigger Level for Inrush: 5.0 A
Trigger Level for Hz Filter OFF : 2.5 A
Trigger Level for Hz Filter ON : 2.5 A

Range : 2000 A; 1400 ac rms

 Resolution: 1 A 
Accuracy > 10 A : 3.5 % rdg + 5 digits
Trigger Level for Inrush: 5 A
Trigger Level for Hz Filter OFF : 8 A
Trigger Level for Hz Filter ON : 8 A

 

Voltage measurement (355 only) dc and ac 10 Hz to 100 Hz 
(600 V and 1000 V ranges have 10 % over range to 660 V and 1100 V respectively.)

Range : 4 V

 Resolution: 1 mV 
Accuracy: 1 % rdg + 10 digits
Trigger Level for Hz Filter OFF: 0.050 V
Trigger Level for Hz Filter ON: 0.050 V

Range : 40 V

 Resolution: 10 mV 
Accuracy: 1 % rdg + 5 digits
Trigger Level for Hz Filter OFF: 0.25 V
Trigger Level for Hz Filter ON: 0.25 V

Range : 400 V

 Resolution: 100 mV 
Accuracy: 1 % rdg + 5 digits
Trigger Level for Hz Filter OFF: 6 V
Trigger Level for Hz Filter ON: 6 V

Range : 600 V ac rms

 Resolution: 1 V 
Accuracy: 1 % rdg + 5 digits
Trigger Level for Hz Filter OFF: 6 V
Trigger Level for Hz Filter ON: 6 V

Range : 1000 V dc

 Resolution: 1 V 
Accuracy: 1 % rdg + 5 digits

 

Voltage measurement (355 only) ac 100.1 Hz to 1 kHz 
(600 V and 1000 V ranges have 10 % over range to 660 V and 1100 V respectively.)

Range: 4 V

 Resolution: 1 mV
Accuracy: 3 % rdg + 10 digits
Trigger Level for Hz Filter OFF: 0.050 V
Trigger Level for Hz Filter ON: 0.050 V

Range: 40 V

 Resolution: 10 mV
Accuracy: 3 % rdg + 5 digits
Trigger Level for Hz Filter OFF: 0.25 V
Trigger Level for Hz Filter ON: 0.25 V

Range: 400 V

 Resolution: 100 mV
Accuracy: 3 % rdg + 5 digits
Trigger Level for Hz Filter OFF: 6 V
Trigger Level for Hz Filter ON: 6 V

Range: 600 V ac rms

 Resolution: 1 V
Accuracy: 3 % rdg + 5 digits
Trigger Level for Hz Filter OFF: 6 V
Trigger Level for Hz Filter ON: 6 V

 

Ohms measurement (355 only)

Range: 400 ?

 Resolution: 0.1 ?
Accuracy: 1.5 % + 5 digits

Range: 4 k?

 Resolution: 1 ?
Accuracy: 1.5 % + 5 digits

Range: 40 k?

 Resolution: 10 ?
Accuracy: 1.5 % + 5 digits

Range: 400 k?

 Resolution: 100 ?
Accuracy: 1.5 % + 5 digits

 

Continuity beeper (355 only)

 

 On at ≤ 30 ?

 

 Off at ≥ 100 ?

 

Frequency measurement

Measurement range

 5.0 Hz to 1 kHz

Resolution

 0.1 Hz (15 Hz to 399.9 Hz); 1 Hz (400 Hz to 1 kHz)

Accuracy – 5.0 Hz to 100 Hz

 0.2 % + 2 counts

Accuracy – 100.1 Hz to 1 kHz

 0.5 % + 5 counts

Trigger level

 Refer to current and voltage tables

 


*Add 2 % to error spec for CF > 2

General specifications

Batteries

 

 Six 1.5 V AA NEDA 15 A or IEC LR6

 

Battery life (with typical usage, backlight off)

 

 100 hours

 

Test leads

 

 Rated to 1000 V

 

Weight

 

 .814 kg (1.8 lb)

 

Jaw size

 

 58 mm (2.28 in)

 

Dimensions (LxWxD)

 

 300 mm x 98 mm x 52 mm (12 in x 3.75 in x 2 in)

 

Safety rating

 

 IEC 61010-2-032, 600 V CAT IV, 1000 V CAT III

 

 

Environmental specifications

Operating temperature

 

 32 °F to + 122 °F (0 °C to +50 °C)

 

Storage temperature

 

 -4 °F to 140 °F (-20 °C to +60 °C)

 

Operating humidity

 

 0 to 95 % (non-condensing)

 

Operating altitude

 

 2000 m

 

Storage altitude

 

 10,000 m

 

IP rating

 

 42 (indoor use only)

 

Drop test requirements

 

 1 m

 

EMI, RFI, EMC

 

 FCC part 15, IEC/EN 61326-1:1997 class B, IEC/EN 61326:1997 3V/m, performance criteria B, EN61325

 

Temperature coefficients

Current:

 0.1 % of reading per °C outside 22 °C to 24 °C

Voltage:

 0.1 % of reading per °C outside 22 °C to 24 °C

 

 